Kategória Kalibračné systémy |
Výrobca GE General Eastern |
|---|

The Humilab Humidity Control Chamber is a NIST traceable relative humidity calibration system that is monitored and precision controlled by a primary chilled mirror hygrometer and RTD temperature sensor.
stability and facilitates cost-effective calibrations of relative humidity sensors, probes, transmitters, recorders and data loggers.
Features
- Primary Standard: Integrated Chilled Mirror & Platinum RTD Constantly Measure Test Conditions
- Precise: ±0.2% RH Stability
- Accurate: ±1% RH from 10-70% RH & ±1.5% RH from 70-90% RH
- Versatile: 10 to 90% RH & 20 to 30ºC
- Self-Contained Operation: Equipped with Desiccant and Water Reservoir
- Connects to External Temperature Bath/Circulator to Standardize Test Conditions
- Connects to External Dry Compressed Air to Extend Desiccant Life Indefinitely
- Record & Document: 0 to 5 VDC, 4 to 20mA Analog Outputs & RS-232 Digital Data Stream
- Fast Response: 10 Minutes for 63% Step Change & 30 Minutes to Full Stability
- Quality Design: Stainless Steel Chamber
- Instant on Fittings for External Dry Air, Circulation of Temperature Controlled Water & Reservoir Drain
- Cost Effective: Ability to Calibrate Several % RH Instruments at Once
- Ease of Operation: Set to Value You Need by the Front Panel or via PROSTEP Software for Ramp & Soak Profiles
- Bright Easy to Read Control Panel Transportable: Carrying Case Supplied